Sam's Laser FAQ, Copyright © 1994-2005, Samuel M. Goldwasser, All Rights Reserved.
I may be contacted via the Sci.Electronics.Repair FAQ Email Links Page.

  • Back to Sam's Laser FAQ Table of Contents.

    HeNe Laser Testing, Adjustment, Repair

    Sub-Table of Contents



  • Back to Sam's Laser FAQ Table of Contents.
  • Back to HeNe Laser Testing, Adjustment, Repair Sub-Table of Contents.

    HeNe Problems and Testing

    How Can I Tell if My Tube is Good?

    A variety of faults can result in a HeNe tube not working properly. These can be the result of a large number of problems including a bad tube, a bad power supply, incorrect input voltage (or configuration), incompatibility between the tube and power supply, bad or incorrect ballast resistor, or defective wiring.

    If all you have is a HeNe tube but no power supply, see the section: Testing a HeNe Laser Tube Without a Compatible Power Supply for ways to determine if the tube is good. The following applies to both bare HeNe tubes and laser heads though some of the inspection and/or tests will require removing the tube from any enclosure.

    Several types of problems can prevent a HeNe tube from lasing properly or make it hard to start:

    It probably doesn't make sense to spend a lot of effort, time, or money to revive a 1 mW HeNe tube that can be replaced for $15. However, if you are ambitious or a new tube cannot be substituted easily (e.g., due to mounting restrictions), see the sections starting with: Repairing Leaky or Broken HeNe Tubes.

    Testing a HeNe Laser Tube Without a Compatible Power Supply

    First time laser enthusiasts are often confronted with this problem: Becoming the proud owner of a HeNe laser tube but no HeNe laser power supply and possibly no specs for the tube or assurance that it is even good. What to do? How can a quick inexpensive test be done to determine if the tube works?

    First, you need to determine the tube's power connections. See the section: Identifying Connections to Unmarked HeNe Tube or Laser Head if you aren't sure.

    There are many ways to power a HeNe tube for the purposes of seeing if it produces a beam. Almost anything that can provide enough voltage to get a few mA through the tube will result in at least a momentary flash of laser light out the end if the tube is good. There won't be any way of determining output power or whether the tube meets specs, but the knowledge that it lases at all may be enough to take the next step - the purchase or construction of a proper power supply.

    It is easy to use the family microwave to see if the tube is gas-intact if the tube will fit inside. See the section: Using a Microwave Oven to Evaluate and Revive HeNe Laser Tubes. While this won't tell you if the tube lases, if it fails this test, there is no need to go further.

    To test for lasing, current must be passed through the bore of the tube. A couple of options for a quick test power supply are:

    Even a high voltage AC supply with appropriate current limiting can be used safely for a few seconds only. (I've been sent HeNe laser tubes which have been operated on AC because the owner copied some power supply design off the Web and didn't know any better. The output power (what of it there was) gradually declined over a few minutes and then there was none.) And even with the rectifier voltage, the tube will be restarting once per cycle which is hard on it so don't run that for too long either. None of these are suitable to operate a HeNe tube continuously unless proper filtering and starting circuitry is added to turn it into a proper HeNe laser power supply.

    Don't go overboard though: Too high a voltage applied in the wrong place can arc straight though the glass at which point you have a rather boring high-tech sculpture. :( A very high current can also damage the tube very quickly, thus the need for the current limiting ballast resistance.

    With these power supplies driving the tube, if there is any output beam, even if it is weak or in the form of short flashes, the tube is probably good. However, there is no way to tell if it meets specs since HeNe laser output power is only maximum over a narrow range of tube current and these quick test power supplies are at most controlling only average current, not instantaneous current as would be the case with a real HeNe power supply. But, at least you know the tube isn't dead.

    Quick and Dirty Universal HeNe Laser Test Power Supply

    The following simple device can be used to confirm that a HeNe laser tube or head is not totally dead. It won't be able to help with output power or be safe to run for more than a few seconds but should produce some coherent output from any HeNe laser from 0.5 to 40 mW or more.

    It consists of the following components in series:

    Wire the output of the transformer in series with the rectifier(s) and ballast resistors. The positive output goes to the anode of the HeNe head or tube; the negative to the cathode. It doesn't matter whether the laser has an internal ballast resistor. Insulate everything VERY well. :)

    Powering the laser should result in flashes of coherent light, probably at the power line frequency (60 or 50 Hz). The amount of light will not be that impressive even with a perfectly good high power laser since the current is nowhere near optimal for any length of time, if ever. However, the presence of laser output would confirm that there is life.

    WARNING: Since centertap of transformer secondary should be grounded, both outputs of the power supply will be floating with respect to ground. Take care.

    HeNe Tube Lases but Color of Discharge Changes Along Length of Bore

    I've only seen this on a couple of HeNe tubes but the appearance is quite strange. At the anode-end of one, which I call the "Northern Lights" tube, the discharge color is perfectly normal - white-ish red-orange (salmon color), just as would be expected in any self respecting HeNe tube. However moving toward the cathode-end of the tube, the color changes to what appears to be a much more blue-ish white color (relative to the normal white-ish red-orange, though in absolute terms it may be more of a pink), typical of some sort of contamination. However, having a tube with the behavior provided a way (as you shall see below) of relating power output to the color of its discharge.

    In this condition, the tube still lased at a power level which relative to its rated output, is approximately proportional to how much of the bore has the correct color. In this sample, about 2 mW for a tube specified at 4 to 5 mW. I don't believe the starting or operating voltage has been affected very much.

    The explanation that makes the most sense is that due to the discharge current in the bore, the few N2 and O2 atoms (and any other party poopers that may have entered without an invitation) are being ionized and pushed toward the cathode of the tube leaving the desired helium and neon atoms to play at the anode-end. The contamination, whether due to a manufacturing problem or an air leak, is so marginal that nearly all of the unwanted atoms are swept from about half the length of the bore. However, the other HeNe tube I have like this had the color change in the exact opposite direct - correct at the cathode but blue-ish-pink at the anode, also reduced power. I now suspect that it may have been internal contamination. More research is needed. :)

    Another unusual characteristic of the Northern Lights tube was that the output power (what of it there is) peaked at a current somewhat higher than expected (8 mA as opposed to the 6 or 6.5 mA typical of this size tube). I don't know whether this is simply due to the overall contamination or that more of the nasty unwanted ions being swept from the bore when running at a higher current.

    This tube had an unfired getter which provided a means of cleaning up the contamination without a refill. A few weeks later, I got around to making the attempt. And the results are.... See the section: Repairing the Northern Lights Tube.

    Starting Problems and Hard-to-Start Tubes

    Some tubes seem to practically start on their own. Other won't perform even when you stand on your head, hold your breath, and provide the proper chants and sacrifices. :-) Note that this may not be a problem under your control. Apparently, even the Ph.D. physics types at major laser companies may not understand why apparently identical HeNe tubes from the same production run may vary in their ability to start by a large amount. See the section: About Hard Start HeNe Tubes. HeNe tube itself isn't to blame.

    However, it could also be that your power supply operating voltage, ballast resistor, and other factors may need modification. Of course, if the system used to work reliably and suddenly died, an actual power supply or wiring problem is most likely though a dead HeNe tube is also possible especially if the system has been unused for several years. But don't overlook the unlikely, but not impossible situation where your line voltage is low for some reason! Check it first. The discussion below is somewhat oriented to the situation where a HeNe tube or laser head is being assembled with a power supply (or parts have been replaced) and the combination just doesn't want to work properly. However, some of it also applies to actual failures as well. Where the power supply itself is suspect, see the section: Power Supply Measurements, Testing, Repair.

    There are several types of possible behavior depending on how well the power supply, ballast resistance, and HeNe tube are matched up, and if any of these as well as the wiring, are faulty. You first need to determine if the discharge is being initiated at all. If the starting voltage is adequate, there will be momentary flashes that may be extremely short and weak and only visible in a darkened room but operating current may not actually follow. Under marginal conditions, operating current will flow in response to the starting voltage but won't be maintained. These flashes will be brighter and longer in duration. The result may be a nice flashing laser. In fact, this progression is exactly what will be seen when operating a HeNe laser tube from a power supply on a Variac as the voltage is increased: Short flashes followed by longer flashes and at some point, a steady beam.

    WARNING: If your HeNe tube doesn't start after a reasonable length of time (like a minute), don't leave the power supply on overnight in a futile attempt to get it going. Starting is a stressful time for power supply components, especially some wide compliance designs, and an extended period with the very high starting voltage on parts of the circuitry may result in total failure. It could also result in electrical breakdown (arcing) inside the laser head or cable. If the laser is flashing, this may be ultimately bad for the tube as well. Turn it off, step back, and try to determine what is wrong.

    Where the power supply components and/or wiring is exposed and subject to dirt and grime, first, carefully clean everything to eliminate possible sources of electrical leakage, which can affect operation, particularly the very low current starting circuit. As an experiment, try warming up the unit (which drives off conductive moisture) with a hair dryer or heat gun on the 'low heat' setting. This may enable it to start more easily confirming the need for some housekeeping. :)

    First, vacuum and/or dust it off with a soft brush, then use mild detergent and water followed by isopropyl alcohol (rubbing or medicinal is fine as long as there are no additives). Give it ample time to dry completely. The hair dryer or heat gun can be used to help it along. You may now find that your starting problems have disappeared!

    If your tube or head has an external starting loop or tape (see the section: Power Requirements for HeNe Lasers), it must be cleaned thoroughly as well (or maybe it has become disconnected, is broken, or has shorted to the case!).

    There is also a possibility that something else is shorting out the power supply, possibly only when enough voltage is applied so it won't show up with an ohmmeter test. Sometimes, the ballast resistor inside cylindrical laser heads will arc to the case. This can be checked with an HV insulation tester or more easily for most people, by removing the end-cap(s) and visually inspecting (as well as smelling!) for evidence of arcing, or by disconnecting the anode wire and driving the tube directly from the power supply with an external ballast resistor.

    Assuming none of this helps, there are three types of behavior: (1) No action of any kind, (2) an occasional flash possibly at random intervals, and (3) a periodic flashing laser which never settles down to normal steady operation. However, the behaviors and their causes are not really always independent so read through all of the possibilities before replacing components or ripping your system apart!

    1. Tube does not fire at all. This means that the tube itself appears totally dead with no flashes inside and no evidence of a beam, even for an instant.

      This generally means that the starting voltage is inadequate for the tube or isn't reaching it, there are other circuit problems, or the tube is bad. In rare cases, shining a light into the tube will allow it to start. Tubes with longer and narrower bores (capillaries) will generally require greater starting voltage and your power supply may just not be up to the task. While tube manufacturers generally specify a starting voltage of 7 to 10 kV (or higher), typical tubes will fire with 3 to 5 times their operating voltage. Thus, a tube that runs on 1,700 VDC will probably start on 5,400 to 8,500 VDC.

      • There may be too much leakage in the anode circuit preventing the buildup of adequate starting voltage. The problem may be in the power supply itself or in the wiring to the HeNe tube. Corona discharge or arcing can result from inadequate insulation or component spacing as well as sharp points in the wiring or connections. Dirt and grime may also reduce the insulation resistance. A sizzling and/or ticking sound along with the aroma of ozone are indications of this sort of HV breakdown. Highly humid conditions may make the situation worse. For pulse (trigger) type starters, there may be too much capacitance as well.

        In the case of an enclosed laser head with a HV (e.g, Alden) connector, HV cable, and internal (potted) ballast resistor, there may be a breakdown in one of these components and it may only show up when starting voltage is applied (not with an ohmmeter). Here are two ways of testing for this situation:

        • Disconnect the anode of the HeNe tube and substitute your own ballast resistor and wiring.

        • Remove the negative connection from the ballast resistor assembly entirely - float it so the starting voltage cannot arc to anything. Connect the negative directly to the cathode of the HeNe tube or laser head case as appropriate.

        If the tube now starts, one of the original components was faulty (most likely the potted ballast resistor assembly if the negative connection runs through it) and this will need to be replaced.

      • If you need to increase the input to start or obtain any sort of response but then must back it off substantially to reduce the tube current to the proper value, low starting voltage or one of the other related problems is indicated.

        Assuming the power supply and wiring check out and the tube is good, the only solution is to boost the starting voltage or use a different type of starting circuit (inverter instead of voltage multiplier, for example).

      • You may have an extremely hard-to-start tube. Whether this is just normal for your particular model tube or this particular sample, is due to it being old or unused for a long time, it is just tired with many hours under its belt, or some other problem, the result is that the specified starting voltage does not have any effect.

        Note that newly manufactured tubes requiring more than a second or so to start using a compatible power supply are usually rejected as defective and may end up in the hands of surplus dealers who may sell them as 'new' even though they don't meet specs. Thus, you may be more likely to end up with one of these hard starting tubes!

        • If a particular tube doesn't start right away because the tube itself is hard starting, cycle the power supply on and off a few times at about 1 second intervals. If it's going to start, it usually will by the third or forth try. I suppose the rapid rate of rise of the starting voltage does something useful. :) (How well this works will depend on the actual tube and power supply but most power supplies should survive the abuse.) While probably not a long term solution, if the tube starts, at least you know it isn't dead.

        • If the voltage doesn't discharge in a reasonable amount of time after powering down due to a lack of bleeder resistor, it may help to discharge the capacitance of the tube and power supply manually with a well insulated high value resistor (a few hundred k ohms to a few M ohms) before attempting to restart. Again, the rapid rate of rise of the starting voltage helps to ionize the gas. Where the tube starts reliably when this is done, it may be possible to add a high value (e.g., several hundred M ohm) resistor permanently across the power supply output.

        • For HeNe laser tubes with exposed bores like Spectra-Physics side-arm tubes, touching various glass parts of the tube (NOT the high voltage connections!) might also provide a capacitive path for the starting pulse. This applies more to pulse starters or those with s substantial component of high voltage AC, not the typical voltage multiplier with a DC output. Older Spectra-Physics exciters tend to have up to 50 percent of the starting voltage in the form of an AC waveform. Double check that the laser head case is grounded - a missing ground may result in starting problems. (More on this below.)

        • Shine a light on the bore in various places to see if that has any effect. On high mileage (probably) tubes, as with high mileage neon lamps or fluorescent lamp starters, a few photons can help ionize the gas. This isn't that common but is easy enough to check.

        • Another test is to try the tube with reverse polarity on its input. Connect the positive output of the power supply to the ballast resistor (don't omit this!) and then to the cathode (can electrode) end of the HeNe tube. Connect the negative of the power supply to the anode of the tube. You are only doing this for testing! Do not be tempted to leave the tube wired this way permanently should it actually start.

          Based on tubes I have tested, the starting voltage is much lower with the anode and cathode connections interchanged. However, the voltage drop across the tube when running with reverse polarity is much higher than with correct polarity. Thus, the tube may not run within the normal operating voltage range of your power supply even if the discharge is initiated - more likely it will just pulse.

          Nonetheless, even if it just pulses, at least you know the tube is not totally dead. If the tube is otherwise undamaged, there should also be an indication of (at least weak) laser output from the business end of the tube. Perhaps, all you need is a power supply with higher starting and/or operating voltage. An inverter type starter using a flyback transformer appears to be particularly good for hard-to-start tubes. Unfortunately, I do not know of any reliable way of determining the likelihood of success without actually trying it.

          I have one 5 mW HeNe tube that requires (depending on its mood) as much as 15 to 20 kV to start (it should be less than about 10 kV). However, once started, it runs with a normal operating voltage of about 1,800 VDC.

          WARNING: Do not let the HeNe tube run for any length of time with reverse polarity as damage may occur due to heating and sputtering at the anode end of the tube.

        • Carefully heating the tube (with a hair dryer, NOT a propane torch!) may help to get it started in some cases. I've only heard of one instance of this and have not tried it myself but it should be easy to experiment. A tube that responds to this treatment is probably one what just has trouble starting the first time after being off for awhile but restarts easily when warm. However, what may really be going on is that you are heating the power supply circuit board nearby reducing leakage in the anode circuit (see above) or bringing a ground in close proximity to the tube (see below).

    2. Tube flashes occasionally for the very briefest of instant, possibly at random intervals. There may or may not be a laser beam accompanying the flashes. Depending on the situation, it may just appear to give up and revert to (1), above. This may be a protective feature of certain power supplies or an indication that the starting pulse is being diverted by an electrical leakage path, corona, or arcing somewhere.

      This sort of behavior is probably more likely with a pulse type starter but can occur with other types as well. What is likely happening is that the energy is insufficient to fully ionize the gas inside the bore of the HeNe tube so the discharge doesn't 'catch'.

      In addition to the other possibilities listed above and below:

      • Check for missing grounds. Where a pulse starter is used or a marginally compatible power supply where the high rate of rise of the starting voltage is needed to start it is possible that like many fluorescent lamp fixtures, a grounded metal plate is needed near or in contact with the tube to provide a capacitive path to aid initial ionization of the gas. Operating the laser (or fixture) on an ungrounded outlet, not installing a foil shield or not properly grounding it, or replacing the HeNe tube with one that doesn't have such a 'feature', might result in erratic starting - perhaps no action at all or maybe an occasional flash but not actually remaining on or taking a long and random time to 'catch'. If putting your hand on the tube (carefully avoiding the high voltage terminals!) results in consistent starting, lack of proper grounding would be confirmed. (This applies to uncooperative fluorescent fixtures as well.) Add a wrap of aluminum or copper foil and attach it to earth ground. I have only heard of one case that might have been due to a missing ground so I suspect this is quite an unlikely scenario!

      • Make sure the anode leads are as short as possible to minimize stray capacitance.

    3. Tube flashes momentarily - longer than an instant possibly up to a second or two - but does not 'catch'. During the time the discharge is on, there is a laser beam.

      What happens is that the discharge is initiated but the voltage drops too much at the tube anode and the discharge goes out. This cycle repeats resulting in a flashing HeNe laser.

      To produce a stable discharge, the following must be satisfied:

      • The sum of the effective resistance of the power supply and the ballast resistor and the (incremental) negative resistance of the tube (dV/dI at the operating point) must be greater than 0.

      • The voltage across the tube must be above the minimum for the tube at the operating current.

      • The current must be above the minimum for the tube/power supply/ballast resistor combination.

      These factors are not independent. Since the negative resistance and sustaining voltage of the tube are not normally specified and depend on current, some amount of trial and error may be required to achieve consistent stable operation but in most cases it really is very easy.

      Cycling behavior can be due to several factors:

      • Poor power supply voltage regulation or excessive ripple. Until the tube fires, there is essentially no load on the supply resulting in much greater voltage than under load. Except for a high compliance type of design where this is needed to produce the starting voltage, minimizing this difference will improve stability and reduce the voltage needed for stable operation.

        If the transformer or inverter drops too much under load, the tube voltage may fall below the minimum for the tube/ballast combination as soon as it starts. This cycle will repeat continuously or it occasionally may catch.

        Use a higher voltage and larger ballast resistor, and/or increase the uF value of the main filter capacitor (and/or the one in the DC supply to an inverter type supply as well if it isn't regulated).

        Minimum capacitor values for less than 5 percent voltage ripple (typical voltage and current requirements):

        • Line operated supplies: .5 to 1 uF (2000 V, 5 mA).
        • Inverter output: .005 to .01 uF (10 kHz, 1,800 V, 4 mA).
        • Unregulated inverter input: 15,000 to 20,000 uF (12 V, 1 A).

        Actual ripple in the current to the tube may be several times greater than this since it depends on the change in voltage with respect to the total effective resistance of the PS+tube+ballast resistor combination). However, the resulting ripple in the optical output power will be 2 to 10 times lower than the ripple in the current depending on operating point. The lowest will occur around the tube's optimal current specification.

      • Ballast resistor too large for the operating voltage. The operating current falls too low resulting in increased (magnitude of) negative resistance. Once the total system resistance goes negative, the discharge becomes unstable and goes out. The result is a flashing laser like a neon bulb relaxation oscillator.

        For an unregulated power supply, increase the operating voltage and/or decrease the ballast resistance.

        For a regulated power supply, decrease the ballast resistance so that the voltage for the desired operating current falls within its compliance range.

      • Too much stray capacitance and/or inductance in anode circuit. The system is behaving like a relaxation oscillator as the capacitance charges and then discharges through the tube. The wiring inductance causes the current from the main supply to lag too far behind the starting current and the discharge goes out.

        Shorten the wiring - minimize the distance between the power supply and ballast resistor, the ballast resistor, and tube anode, and don't use long runs of high voltage coax (which may have higher capacitance). Increasing the energy of the starting circuit slightly may help as well.

      • For laser heads in particular, the additional capacitance resulting from the metal case may increase the minimum stable tube current by up to 1 mA or more - and thus require changes in the power supply and/or ballast resistor. So, if you tested the HeNe tube and power supply on your workbench but the enclosed system is unstable, this may be the reason.

      • Power supply polarity is reversed. The voltage drop across a HeNe tube operated with the cathode and anode interchanged is higher than under normal conditions. However, required starting voltage is much lower. The result is likely to be a pulsing laser. Double check your wiring and terminal connections. I have also seen commercial power supplies mislabeled! See the section Making Measurements on HeNe Laser Power supplies if you need to actually test for reverse polarity.
    Also see the sections: How Can I Tell if My Tube is Good?, About Hard Start HeNe Tubes, Testing a HeNe Laser Power Supply, and Power Supply Construction Considerations.

    About Hard Start HeNe Tubes

    As noted elsewhere, apparently identical HeNe tubes made on the same production line may differ widely in their starting performance. While one sample may start absolutely instantly, another one that is indistinguishable may take several minutes to wink on. A newly manufactured HeNe tube that requires more than a second or so to start using a compatible power supply is generally rejected as defective. Guess who may end up with these - surplus dealers, and ultimately, you! The condition may get worse with use so some high mileage tubes could indeed be a lot worse in this department - possibly to the point of being virtually impossible to start even if they were within spec when new.

    As far as I can determine, the fundamental physics behind this phenomenon may not even be well understood by the major laser companies. The only meaningful data is statistical, because even a give tube with a given power supply will have dramatically different start times from attempt to attempt, as will tubes built side-by-side through the entire production process.

    Tubes that are kept in dark cold environments for long periods of time don't tend to start well. But, once one of these tubes is started successfully, restarts will likely be instantaneous, or at least reasonably quick. However, left overnight, they will revert to being uncooperative.

    Also lower fill pressures and cleaner tubes make for hard starting - not to mention power supply variables.

    Some manufacturers (e.g., Melles Griot) use a conductive 'start-tape' running the length of the tube attached to the anode electrode to aid in starting. It's not even really proven that this improves performance (and I've found that it can be a source of electrical breakdown problems. I've never noticed any difference in the speed of starting after removing the start-tape). Uniphase had a pointed electrode inside the anode mirror sleeve to aid in starting but it isn't obvious that it made any statistical difference either. There has even been talk of using a trace of radioactive gas (as used to be common in neon indicator lamps and glow tube fluorescent starters), but this of course would probably not be a popular idea today!

    A given production line may still have hard-start related yield problems from time to time (which kind of suggests the Ph.D physicists don't understand it). Funny thing is, no one can tell anything that's different on a hard-starter versus a regular one.

    Possible Causes of No Output with a Normal Discharge Glow

    Where the HeNe tube starts - there is a stable glow discharge and it is the correct color (bright white-ish red-orange, see the section: How Can I Tell if My Tube is Good?), there are a few possibilities that are not due to a bad HeNe tube or laser head:

    And, for other-color HeNe tubes which have much lower gain for a given length than red HeNes, all of the above may apply. The following comments were prompted by questions about a non-lasing short green HeNe tube (similar to a Melles Griot 05-LGR-024, 215 mm in length:

    (From: Lynn Strickland (stricks760@earthlink.net).)

    Those things are touchy, touchy, little SOBs. They usually have an almost flat HR and OC combination. If it does lase, it will probably be a few tenths of a mW at best. Probably have to walk the beam AND tweak both ends for any hope. Try some magnets too, for 3.39 micron suppression. In general, low power greens are a bitch to tweak.

    Note that the green discharge is more 'pink' (red tubes more 'orange'). Fill mixture is a little different, but the different color mostly due to lower fill pressure - which is why greens have shorter lifetimes than red.

    Possible Causes of Low Output with a Normal Discharge Glow

    Where the discharge color is normal and the tube current is known to be reasonably correct but the output is weak, there can be a variety of causes. (Of course, most of this also applies to no output but this is covered in the section: HeNe Tube Problems and Testing.) The following applies when the output power just isn't as great as expected:

    Doughnut Mode Beam from TEM00 HeNe Laser

    Occasionally, a HeNe (or other) laser designated as TEM00 will actually output a beam which looks like a doughnut or torus - with a hole in the middle. It's running with a not quite TEM00 mode structure - a doughnut mode. Since this should be a TEM00 laser with a single transverse mode Gaussian beam profile, that's normally considered a bug. If it's outputting decent power (compared to its specifications), the cause isn't likely to be dirt inside the tube or imperfect mirror alignment.

    More likely, the manufacturer accidentally used too large a bore for the length of the resonator and the mirror curvature. For example, if this is a green (543.5 nm) HeNe laser, they may have used a bore sized for a red (632.8 nm) HeNe laser by mistake resulting in a mode diameter that is too large. Or, it might have been designed on the hairy edge, size-wise, in an attempt to get as much power as possible out of the tube and the engineers weren't lucky that day.

    If you had been the original owner, the laser might have been replaced under warranty. As it is, you now have what I generally call an "interesting" laser. :) Or, since the specification are often only with respect to "95% mode purity", if the hole represents less than 5 percent of the power, maybe it's considered acceptable, though I can't imagine anyone being entirely happy with a laser that's supposed to be TEM00 having a hole in the middle of the beam!

    In any case, enjoy the unusual behavior. There's nothing you can do about it!

    Unstable or Flickering HeNe Tube

    Where your HeNe tube starts and lases normally but is unstable, flickering or going out and then restarting whenever the power line voltage dips slightly or for no apparent reason, the problem may be power supply or HeNe tube related. It is also not unusual for this to start happening after the system has been on for awhile due to characteristics of one or more of the components changing slightly.

    A different power supply or slight adjustments or modifications may make your HeNe tube happy, at least temporarily. However, where the HeNe tube is an inexpensive vanilla flavored variety, replacement may be the easiest solution if it turns out to be marginal. :-)

    Sputtering or Erratic HeNe Tube

    This behavior is somewhat similar to the some of the compatibility problems described above. However, it could be much more random and may only occur during warmup.

    The symptoms are that the tube may start normally but then go off and restart, possibly quickly and unpredictably. One possible cause is a bad internal connection between the cathode can and its attachment to the mirror mount where the negative lead of the power supply is hooked up. The type of construction susceptible to this malady is where a 'nipple' on the end of the cathode can is swaged (pressed/squished) into the mirror mount rather than actually being attached by spot welding or via a spring contact. After many thermal cycles, the swage can loosen resulting in intermittent contact especially as the tube heats and parts expand. Any sort of high resistance increases the required tube voltage since the mirror mount has a higher 'cathode fall' voltage drop. The discharge will likely go out and the power supply will then attempt to restart. In some cases, the discharge may strike to the mirror mount itself (look for a glow near the mirror) and if this persists, will eventually destroy the mirror. (See the section: Damage to Mirror Coatings of Internal Mirror Laser Tubes) After the tube warms up sufficiently, since aluminum expands faster than steel or Kovar, the problem may disappear once the connection tightens. However, until then, the intermittent contact and many restarts is hard on the power supply and possibly the mirror.

    Assuming the power supply and tube are properly matched and the power supply isn't defective, this is a defective HeNe tube. No cure is possible. This is a relatively unusual problem (I've only seen it in two (2) HeNe laser tubes so far) so first check external connections and make sure your HeNe tube and power supply are properly matched. If its maximum voltage is marginal, as the tube heats up, the voltage drop may increase just enough to result in erratic behavior. However, one possible difference between this and a bad cathode connection is that with the latter, the condition may clear up once the tube heats up since the expansion of the aluminum cathode will improve contact. The marginal voltage situation will just get worse. The power supply itself could also be defective. The easiest way to determine which is at fault is to swap the PSU and/or tube with known good units.

    Also see the section: Unstable or Flickering HeNe Tube.

    Cyclical Variations in Output Beam Power

    A HeNe laser that is in good condition will produce an output beam that is quite stable and will have no visible (at least by eye) variations in output power though a laser power meter will show fluctuations over various time scales of a few percent even after warmup unless it is high performance amplitude stabilized laser. A typical laser tube will have a "mode sweep" specification of between 2 and 20 percent depending on type and size (smaller tubes typically have poorer performance). As the tube warms up, this cycling will start out slowly, speed up and peak, and then taper off as the system stabilizes thermally.

    Note that if the discharge is actually going on and off, the cause is entirely different - an incompatibility with the power supply, incorrect ballast resistor, low line voltage, etc. See the section: Unstable or Flickering HeNe Tube.

    However, sometimes you will find a laser that exhibits significant periodic variations in output intensity even where the discharge is perfectly stable. There are two types of phenomena depending on the period of the power cycles:

    Reasons for Short HeNe Tube Lifetime

    As noted, sealed HeNe should last many thousands of hours with no noticeable degradation in performance.

    If you are experiencing excessively short life (e.g., a month instead of years), the first things to check are operating current and polarity. See the section: Making Measurements on HeNe Laser Power Supplies. Of course, if you omitted the ballast resistor, life will likely be very short. :-(

    If the HeNe tube and power supply are mismatched, one can damage the other. For example, running a 1 mW HeNe tube on a power supply designed for a 35 mW HeNe tube may not only result in too high a current by design (e.g., 8 mA instead of 3 mA) but may also result in much higher current if the compliance range of the power supply is exceeded (i.e., the voltage across the HeNe tube is much lower than the power supply can handle). Conversely, attempting to power a 5 mW HeNe tube using the power supply from a barcode scanner (designed for a .5 to 1 mW HeNe tube) will likely result in a blown power supply. Just because the high voltage connectors mate and/or the tube lights up doesn't imply anything about compatibility! Also note that maximum optical output occurs at the optimum operating current - too high or too low and it goes down. (Operating current for yellow, orange, and green HeNe tubes is even more critical than for the common red variety so setting these up with an adjustable power supply or adjusting the ballast resistance for maximum output is recommended.)

    New and even used HeNe tubes and power supplies from reputable surplus dealers will generally last a long time if not abused. But, much of what you get at swap meets and hamfests has been pulled from equipment for one reason or another. So, the problems you are experiencing may have nothing to do with your setup!

    (From: Lynn Strickland (stricks760@earthlink.net).)

    Speaking as a non-physicist....

    There are so many variables in a gas discharge, it's a game of averages. That's why the power supply business can be so tricky - and why, for the power supplies you can look inside of, you see so many modifications. That, and the rate at which electronic components go obsolete keeps it in a continuous state of flux (no pun intended).

    Reasons for the variability in lifetime and failure mechanisms from design to design revolve around design fill pressure and gas mix, operating current, distance from capillary bore-end to cathode, optical design (some designs are more sensitive to misalignment than others). Also power supply variability, ballast resistor value differences, operating current tolerances (often set at, say, +/-0.2 mA).

    Gas lasers can be a pain, but for a lot of applications, they're still the most cost effective solution -- in some cases the only solution.

    HeNe Tube Use and Life Expectancy

    You often hear that lasers like to be run to keep them healthy and maximize life.

    For both types of HeNe tubes (as well as other lasers), power and beam quality will peak only after some warmup period. So it makes sense to keep the laser energized continuously over the course of an application where these are critical but this has no bearing on any need to turn the laser on just to keep it healthy.

    Here is a chart of very rough guidelines for evaluating HeNe lasers. This is based solely on my observations with only minimal input from those who should know about this sort of stuff like major laser companies:

     Characteristic         New      Middle Age    High Mileage    End-of-Life
    ----------------------------------------------------------------=-----------
     Operating voltage      Spec        +5%            +10%        +25% or More
     Output power          1.5-3X     1.25-2X         1-1.5X          0-0.5X
     Discharge color       Normal     Normal         More Pink      Pink-White
     Brown crud in bore     None       Some            Much            Much
    

    When new, the tube operating voltage will be close to the spec'd value. As the tube is run, it will increase gradually until end-of-life where it may become dramatically higher. Output power on average may actually increase during the early stages of the tube's life due to initial overfill, but will then decrease gradually until end-of-life, where it will nose dive. Output power may also change as a result of slight mirror misalignment caused by multiple thermal cycles but this is correctable. With use, the gas pressure inside the tube decreases resulting in a trend toward becoming more pink. But at end-of-life, sputtering results in a tendency to become more white or purple. Brown crud develops with use, though not all tubes will have any. In many cases, there will be very distinct striations in the brown crud, indicating plasma oscillations, but this appearance alone may not imply much of anything about the tube's health.

    See the section: Case Study of 145 Melles Griot 05-LHR-640 HeNe Laser Tubes for an example of the behavior of one specific model. I do not know to what extent these data apply to other HeNe lasers.

    Brown Deposits Inside HeNe Tube Bore

    Many surplus HeNe laser tubes - both hard-seal and soft-seal - will have various amounts of a brown material coating portions of the inside of the bore, usually toward the anode-end. The presence of these unsightly deposits has no significant impact on operation or power output directly, but is an indication that the tube has seen a lot of use. And, on average, tubes with a lot of brown crud may be harder to start and require higher voltage to run - but that's more related to the high mileage than the preseance of brown crud in the bore. On high power lasers with IR suppression magnets, the brown stuff will generally collect near the magnets with obvious effects of N and S polarity. I've yet to see an explanation for this phenomenon in any laser reference. One suggestion from someone from a major HeNe laser manufacturer was that it was material sputtered off the anode but the one below makes more sense.

    (From: Chris Leubner (cdleubner@ameritech.net).)

    The usual cause is silicon being freed from the oxygen in the glass due to the intensely hot plasma on it. The ionized oxygen ends up reacting with the getter or cathode leaving elemental silicon film behind causing that brown look. In some tubes it will make a zebra or tiger stripe pattern on the bore that is a dead giveaway of both long use and plasma oscillation. On larger tubes that use magnets for IR suppression (Zeeman splitting), the magnetic fields smash the plasma into the tube wall and increases the rate of dissociation of the glass. The oxygen, which is a gas, will disperse throughout the tube and combine with the more reactive materials in it, namely the getter or cathode. The silicon will remain behind wherever it was separated because it is not volatile and relatively difficult to ionize. I do not know why it appears first on the anode end. My guess is probably due to the larger number of negative ions there reacting with the silica in the glass via this reaction: SiO2+2Ne-1=SiO+O-2+2Ne. Then SiO+2Ne-1=Si+O-2+2Ne.

    Care of HeNe Laser Tubes

    The often quoted party line is that soft-seal HeNe lasers should be run periodically to maintain their health and that this is not necessary for hard-seal tubes. The true story is a bit more complex.

    All modern internal mirror HeNe laser tubes use hard-seal construction where everything but the mirrors (where the required high temperatures would destroy the coatings) use glass-to-metal seals. Mirrors are either sealed with frit (low temperature glass powder which acts as a sort of solder for glass), optical contacting, or are fully enclosed inside the glass envelope. None of these seals leak on any time scale that matters unless the processing was defective. Melles Griot quotes a 12 year shelf life but in reality, it's virtually unlimited.

    Note that frit is quite soft compared to even optical glass so don't unnecessarily abuse the mirror seals. Those with large amounts of frit like Melles Griot and Siemens are fairly robust. But the mirrors on those with only a thin frit line like Aerotech and Uniphase may pop off if whacked the wrong way. Unless your intent was to salvage the mirrors, this would be bad news.

    However, there are still many external mirror HeNe lasers that use soft-seals for the Brewster window(s) and these show up surplus with varying degrees of leakage. Tubes of the same age may differ greatly in their condition. apparently due to large variations in the rate of leakage. Where the discharge color is still a pastel but quite bright - somewhat more pink than normal, even with a bluish tinge - just running the tube for a few hours or days may clean it up irrespective of the condition of the getter because the cathode itself acts as a getter - a very slow one but good enough to scavenge a small amount of contamination. The typical discharge color that is still salvageable would be the "Minor" examples in Color of HeNe Laser Tube Discharge and Gas Fill, perhaps slightly worse. Even a HeNe tube that doesn't lase at all may benefit from this simple treatment. Periodically running soft-sealed HeNe laser tubes without getters or with exhausted getters is recommended. A few hours every month is probably adequate and this will extend their life considerably, possibly indefinitely. This is much preferred compared to restoring power once it's gone. Note that any detectable (by eye) change in discharge color will be accompanied by a significant drop in output power. As the tube is operated, the discharge color will gradually approach the correct one. The last place where a normal color appears will be the expanded regions of tubing (e.g., in the glass tube that joins the side-mounted cathode to the bore in a Spectra-Physics laser). Here, the normal color is a nice orange but will tend toward pink or pinkish-blue with contamination.

    Remarkably, for a soft-seal tube, the bottom of the "Minor" samples may actually be easier to salvage by running for a few hours. I've revived both a very old SP-130B as well as a not quite so old SP-120 using this simple treatment. Both these lasers were discarded because based on the color of the discharge, the original owners thought they were too far gone for there to be any hope. The SP-130B only recovered to about one third its rated power (but it is over 30 years old!). Running it every few days for a couple minutes appears as though it will maintain that power indefinitely. The SP-120 was restored to essentially new specifications.

    However, if the discharge color is highly saturated red or blue (the bottom two examples in the above diagram) and/or there are visible striations of the discharge in the expanded regions of tubing, all hope is probably lost as no amount of operation or getter reactivation will make enough difference to matter. But there is nothing to lose by running the tube for awhile to see if a miracle occurs. :)

    When powering a HeNe tube with an off-color discharge, keep in mind that the operating voltage may be quite different than normal especially initially and may overstress the power supply if it doesn't have enough compliance. A brute force unregulated power supply on a Variac can also be used, adjusting the Variac to maintain a more or less constant current at the rated value for the tube. It's also nice to monitor the laser's output (assuming there is any eventually!) with a laser power meter to keep track of how the patient is responding to treatment. What may happen is that the power will initially increase, then decrease as the tube heats up and internal parts outgas, then gradually decrease again as the cathode acting as a getter scavenges the contaminants, and then level off. This process may take several hours or days. Powering the laser on successive occasions may result in increasing power levels if the process wasn't complete. In any case, it won't hurt to try.

    Hard-seal red (632.8 nm) HeNe tubes generally will not respond to these sorts treatments since there should be essentially no leakage over any time scale that matters. The gain - as modest as it may be - is suffient that any improvement may be detectable only by careful power measurements before and after. But there can be exceptions. I did have a modern Melles Griot internal mirror HeNe tube that had an off-color discharge and low power. Running it for several hours didn't help at all but activating the getter with my Solar furnace rig completely cured it permanently (it's been over two years now with no degradation in discharge color or output power so this tube isn't a "leaker" but must have not have been properly processed at the factory). See the section: Repairing the Northern Lights Tube.

    However, for "other color" HeNe lasers, particularly yellow and green ones which have very low gain (about 1/20th of red), running even a hard-seal tube for a few hours *before* thinking about touching mirror alignment can make the difference between nothing and something, even if that something is small.

    I've found some hard-seal HeNe laser tubes where the gas fill was obviously contaminated on the shelf. One example was the HeNe laser tube from a Hewlett Packard 5501A two-frequency (Zeeman split) laser head that hadn't been used in about 15 years. It wouldn't lase at all when first powered up. After running for a total of about 12 hours, it has recovered probably to essentially normal output. This type tube is of very high quality construction and no doubt was very expensive with glass-to-metal seals for electrical connections and mirrors fully enclosed inside the glass envelope. Leakage is unlikely so it must have been internal outgassing over time. Thus, even hard-seal tubes can suffer from soft-seal maladies! :) After being idle for about 2 years, the power had again declined, but only to about 25 percent of the recovered level. Running for awhile again restored it, with a rapid recovery to about the 50 percent level in a minute or so and back to 100 percent in a few hours.

    Note that end-of-life tubes will often show an off-color discharge which may be mistaken for leakage. Output power will be low or zero and there will often be evidence of shiny metallic sputtering deposits on the glass near the cathode can - a dead give away that the tube is end-of-life. They will also likely be hard to start with a very high operating voltage. On Melles Griot tubes, there will be shiny metallic deposits on the glass opposite the three holes at the rear end of the cathode can. On Hughes-style tubes, it will be on the glass at the cathode end of the tube. These tubes will not respond to any known treament.

    Troubleshooting an External Mirror HeNe Laser

    In addition to all of the problems of internal mirror HeNe tubes, external mirror lasers are subject to dirty optics and much more prone to have misaligned mirrors.

    Like their internal mirror counterparts, the general appearance of the output when non-lasing will be a diffuse blue, blue-green, or purple spot but no red light. If there is any evidence of a red beam, something may be marginal but it is lasing.

    If it won't start, then the tube could be up to air or there could be a power supply problem. Try another power supply if available. Or, see the section: How Can I Tell if My Tube is Good? for info on using a low level RF or microwave source to check for ionization.

    Assuming the tube lights up, follow the steps below to narrow down the cause:

    1. If possible, check the tube current. Though unlikely, a discharge current much much greater than optimum will result in low or no output beam (as well as overheating of the tube, ballast resistor(s), and power supply components.

    2. Compare the color of the discharge in the narrow bore/capillary with Color of HeNe Laser Tube Discharge and Gas Fill. The comments about output apply to red and maybe orange HeNe tubes; yellow and green HeNe tubes will likely produce no output at all unless the gas fill is nearly perfect. The normal appearance is a white-ish red-orange generally described as "salmon color" though there can be a fair range from more orange to more pink for a good tube depending on the exact gas fill He:Ne ratio and pressure. Lower pressure tends toward pink and may be normal for low gain non-red HeNe tubes to boost gain at the expense of tube life. Some of these may also be even more of a white-ish shade. To confirm, check the discharge spectrum, preferably with a spectroscope but a diffraction grating or prism may be adequate. It should be similar to the combination of the helium and neon spectra in Bright Line Spectra of Helium and Neon. If the color inside the bore appears normal, check the color of the discharge where it isn't as constricted - the color should be quite orange. (This can be seen in the funnel area near the anode on most internal mirror tubes or the expanded tubing sections on those like Spectra-Physics side-arm cathode tubes with exposed capillaries.)

      • Where both these colors are correct (salmon and orange), gas fill is probably not the immediate problem.

      • If the discharge color in the expanded areas is the same as the bore or more towards white or blue, the gas fill is somewhat contaminated and marginal - lasing may not occur (definitely not for yellow and green HeNe tubes).

      • A slightly more pinkish discharge in the bore but with the normal orange color elsewhere may indicate low gas pressure and near/at end of life. There is no cure (but as noted above, this may be normal for some tubes).

      Firing the getter (if any) or just running the tube for an extended period of time may clean up any slight contamination (but won't help low gas pressure). However, if it is very pink, blue, purple, or white, a significant amount of air has leaked in over time, probably via the soft-sealed Brewster windows, and the only cure is likely to be a tube transplant. This is probably the most common problem with older external mirror HeNe lasers. Unfortunately, it isn't cost effective to refill them and replacement tubes are likely to be very expensive - if they are available at all.

    3. Check the mirrors, Brewster windows, and any other intra-cavity optics for damage and clean them if necessary using the proper optics cleaning technique.

      See the section: Cleaning of Laser Optics for the recommended procedure.

    4. The only remaining cause of a non-lasing laser with a proper discharge color and clean optics is mirror alignment. If the laser was dropped (Ack!) or someone decided the alignment screws were loose and tightened them, see the section: >Sam's Approach for Aligning an External Mirror Laser with the Mirrors in Place or the more general procedures starting in the section: External Mirror Laser Cleaning and Alignment Techniques.

      Of course, this assumes that the optics are correct for the laser or that someone didn't remove a mirror for use in their science fair project! Note that alignment is super critical, especially for a long HeNe laser. Thus, if misalignment is found to be the problem, it may require a lot of patience, determination, and the proper jigs, to remedy it. You won't succeed by luck alone (though luck may play a part)!

    External Mirror HeNe Laser Health Checklist

    When considering the acquisition of a large-frame HeNe laser, here are some specific things to look at or questions that can be asked of the owner which will help to determine if the laser is likely to be functional near original specifications. The following applies directly to external mirror Spectra-Physics HeNe laser tubes with minor modifications for other manufacturers/models:

    Once the laser can be powered up, check the discharge color in the bore. It should be similar to the bright white-ish red-orange or 'salmon' color at the top of Color of HeNe Laser Tube Discharge and Gas Fill, or of any other fully functional HeNe laser tube. If it does not, either the tube is soft-seal and has leaked, or it has been very totally abused. See the sections starting with: HeNe Tube Use and Life Expectancy. If the discharge color looks good, then very likely mirror alignment is all that is needed to achieve at least a substantial fraction of full power.

    Can I Increase Output Power Using the Waste Beam from the HR?

    When operating a bare HeNe laser tube, you've no doubt noticed the weak beam that exits from the supposedly totally reflecting mirror (the HR or High Reflector). (This assumes it isn't covered with tape or paint.) So, could the power from the output end of the laser be increased by putting another mirror behind the HR?

    The quick answer is that this might be possible in theory.

    The practical answer is: forget it.

    The long answer is too involved to go into here but if the extra mirror were properly aligned AND an exact multiple of 1/2 wavelength of 632.8 nm from the other mirror AND if there were no losses from the non-AR coated HR surfaces, part of the wasted power might appear at the output.

    But, in the end, all you would gain at most would be the couple microwatts that escapes out the HR. :) The lost power isn't much on most tubes. For those occasional tubes where the output is significant from the HR (either because of a mistake in manufacture or by design), there might be more benefit but as a practical matter, there is no way to satisfy all the conditions in a stable manner without a fancy feedback loop, if at all.

    (From: Steve Roberts (osteven@akrobiz.com).)

    Assuming it's a standard TEM00 mode HeNe and not a multimode laser, you'd see little tiny increases and decreases in the power on a very sensitive power meter as the mirror was translated toward and away from the existing rear mirror. But you would not really recover any of the rear beam, in fact you'd confuse the lasing going on inside the main cavity somewhat, and at certain possible "magic" combinations of external reflector and distance, cause lasing to actually cease. In practice, HeNe lasers tend to run by default at their maximum possible gain for a given combination of tube optics.

    If you want to see one wink out or flicker, precisely anchor it to a stable bench and then use a third flat mirror some distance away on a precision mount to reflect the output back down the bore. When the reflected beam is 180 degrees or so out of phase with the wave in the cavity, it will wink and flicker.

    (From: Sam.)

    I wonder about this...

    To actually interfere with lasing in a typical HeNe laser may be more difficult than Steve claims. While flickering and apparent instability will be seen if this experiment is done with a common HeNe tube, it may only be a result of the output beam interfering with itself outside the cavity when reflected back to the OC. This could appear to be confusing lasing but may actually not cause any substantial effect inside the cavity. Monitoring the waste beam (as noted below) can be used to determine whether the behavior is due to external or internal interference. If it's only external, the waste power will be almost unaffected (just the portion of the reflected output beam that gets back through both the OC and HR). This is likely to be less than 0.1 percent of the output power or a couple percent of the waste beam power at most. However, if actual lasing is being affected, the waste beam power will fluctuate significantly - up to (as Steve suggests), total wink-outs. :)

    (From: Bob.)

    On a somewhat related side note, there is at least one commercial instrument I know of that focuses the output of a HeNe laser onto a surface, and has a highly sensitive photodetector behind the HR of the laser (the arrangement Steve mentioned, but in reverse). As the surface the light is focused on moves back and forth in relation to the laser, the photodiode detects changes in output power out the back end. Basically, this is a form of a Fabry-Perot interferometer which can be used to very precisely measure small distances.



  • Back to HeNe Laser Testing, Adjustment, Repair Sub-Table of Contents.

    Connections, Electrical Characteristics, Output Power

    Identifying Connections to Unmarked HeNe Tube or Laser Head

    Where you can visually inspect the wiring, this is trivial as we all know that the large aluminum 'can' electrode is the cathode (negative) terminal.

    CAUTION: While most modern HeNe tubes use the mirror mounts for the high voltage connections, there are exceptions and older tubes may have unusual arrangements where the anode is just a wire fused into the glass and/or the cathode has a terminal separate from the mirror mount at that end of the tube. Take note of the cathode arrangement in particular because the tube will still lase perfectly if you attach to the mirror mount but instead of the actual cathode but that will result in sputtering near the mirror which is about the worst place for this - similar to running the tube on reverse polarity. (Miswiring the anode might result in no or weak lasing but probably no permanent damage.)

    Alden high voltage connector

    The two pin Alden is by far the most common connector used for attaching HeNe laser heads to HeNe laser power supplies. They are used by almost all manufacturers and for lasers almost all sizes. The shorter (narrower) side goes to the anode (positive) and the longer (fatter) side goes to the cathode (negative). When such a connector is present, there will also be a ballast resistor (typically about 75K ohms) built into the HeNe tube assembly or laser head between the Alden's positive terminal and the anode.

                                  ________-__
                     Anode (+) ==|________|  |---_______
                             _____________|  |  |_______ HV Cable   
              Cathode (-) ==|_____________|__|---
                                          -
    

    Or see High Voltage Cable with Male Alden Connector. This one is built with separate wires and appears to have a ballast resistor built into the anode (red) lead (or maybe it's just a wart!). Many use coax similar in thickness to RG58U for the HV cable instead.

    Note: Genuine Alden brand connectors will have the name stamped on the plastic. Some power supplies may come with Alden compatibles without identification. This probably doesn't matter in any way, shape, or form, except as an indication that the power supply manufacturer installed the connector onto existing wiring or saved a few cents. :)

    Three pin in-line high voltage connector

    Some larger HeNe lasers (mostly from Siemens and Spectra-Physics) use a connector somewhat similar to the standard Alden but it is more rectangular with 3 pins instead of 2. And, the pins on both of the connectors (male and female) are recessed to avoid the shocking experience of touching the pins on a recently detached laser head and getting zapped!

    CAUTION: The pinouts for Siemens and Spectra-Physics lasers have the HV pins swapped! Using the wrong one may result in very rapid destruction of laser head and/or power supply, not to mention a possible shocking experience!

      Pin Location     Siemens     Spectra-Physics
     ----------------------------------------------
      Square end     Earth Ground   Earth-Ground
      Middle         Cathode (-)      Anode (+)
      Round end       Anode (+)      Cathode (-)
    

    For these large laser heads, there may also be a small ballast resistor in series with the cathode lead. Bypassing it will reduce operating voltage requirements and the laser will probably still work fine though the claim is that stability will be better with it when used with the recommended power supply.

    Three pin round High voltage connector

    Some Spectra-Physics lasers use a special 3 pin round connector (view is looking toward power supply):

                   O  Positive (Anode)
                   1
          GND  O 3   2 O  Negative (Cathode)
               
                   o  Interlock Prong
    

    The GND may not actually be present on some power supplies. In most cases, it is already connected to the negative elsewhere. The interlock prong activates a microswitch in the power supply to complete the primary-side circuit only if the power supply and laser head are securely attached. This provides protection for the power supply but isn't present on all